User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: PM 5138A (Performance) Ver 1.0 DATE: 05-Jun-94 AUTHOR: User Contributed REVISION: 1.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 86 CONFIGURATION: HP 436A CONFIGURATION: Philips PM 3365A CONFIGURATION: Philips PM 5191 CONFIGURATION: Philips PM 6654C CONFIGURATION: HP 8568A CONFIGURATION: Fluke 8842A CONFIGURATION: HP 8901A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Source: Operating Manual PM 5138A P/N: 4822 872 10113 @940312 * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * TUR Configuration: HP 436A, Philips PM 3365A, Philips PM 5191 * Philips PM 6654C, HP 8568A, Fluke 8842A, HP 8901A * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * When the T.U.R. is < 2:1, the source instrument, e.g. 5700A, must be * characterized and appropriate corrections entered in the correction file * (e.g. 5700.COR). Where T.U.R.'s are > 2:1 but < 4:1, the test tolerance * is reduced by root sum square (rss) techniques. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * This procedure is not executable in MET/CAL versions below ver. 2.00 * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Fluke makes no warranty, expressed or implied, as to the fitness or * useability of this procedure. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * The following subprocedures are called from this main procedure: * Sub PM 5138A: 5.4.1 - 5.4.4.4 RS232 * Sub PM 5138A: 5.4.5 - 5.4.6.1 RS232 * Sub PM 5138A: 5.4.6.2 - 5.4.8.3 RS232 * Sub PM 5138A: 5.4.8.4 - 5.4.10 RS232 * Sub PM 5138A: 5.4.1 - 5.4.4.4 IEEE * Sub PM 5138A: 5.4.5 - 5.4.6.1 IEEE * Sub PM 5138A: 5.4.6.2 - 5.4.8.3 IEEE * Sub PM 5138A: 5.4.8.4 - 5.4.10 IEEE * Sub FM RESIDUAL Measurement * Sub AM Measurement * Sub FM Measurement * Sub THD Measurement * Sub PM 51xx RISETIME * Sub PM 51xx FALLTIME * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ 1.001 ASK- R P F W 1.002 HEAD INTRODUCTION 1.003 DISP This procedure is intended to: 1.003 DISP - Check the instrument's specification. 1.003 DISP - Check the nessesity for re-calibration. 1.003 DISP 1.003 DISP NOTE: UUT Requires 30 Minutes warmup before testing. 1.003 DISP 1.004 DISP If the UUT does not have a remote interface, install 1.004 DISP the PM513x IEEE test interface. 1.004 DISP 1.005 OPBR Does the UUT have a IEEE option. 1.006 JMPT 1.013 1.007 DISP Set the UUT RS232 interface for: 1.007 DISP Baud: 9600 1.007 DISP Parity: None 1.007 DISP Bits: 8 1.007 DISP Parity: 1 1.007 DISP XON/XOFF: Off 1.008 CALL Sub PM 5138A: 5.4.1 - 5.4.4.4 RS232 1.009 CALL Sub PM 5138A: 5.4.5 - 5.4.6.1 RS232 1.010 CALL Sub PM 5138A: 5.4.6.2 - 5.4.8.3 RS232 1.011 CALL Sub PM 5138A: 5.4.8.4 - 5.4.10 RS232 1.012 END 1.013 CALL Sub PM 5138A: 5.4.1 - 5.4.4.4 IEEE 1.014 CALL Sub PM 5138A: 5.4.5 - 5.4.6.1 IEEE 1.015 CALL Sub PM 5138A: 5.4.6.2 - 5.4.8.3 IEEE 1.016 CALL Sub PM 5138A: 5.4.8.4 - 5.4.10 IEEE 1.017 END 1.018 IEEE [@436] 1.019 IEEE [@3365] 1.020 IEEE [@5191] 1.021 IEEE [@6654] 1.022 IEEE [@8568] 1.023 IEEE [@8842] 1.024 IEEE [@8901]